Article Dans Une Revue Clinical Hemorheology and Microcirculation Année : 2018

Indirect viscosimetric method is less accurate than ektacytometry for the measurement of red blood cell deformability.

Résumé

The aim of this study was to test the accuracy of viscosimetric method to estimate the red blood cell (RBC) deformability properties. Thirty-three subjects were enrolled in this study: 6 healthy subjects (AA), 11 patients with sickle cell-hemoglobin C disease (SC) and 16 patients with sickle cell anemia (SS). Two methods were used to assess RBC deformability: 1) indirect viscosimetric method and 2) ektacytometry. The indirect viscosimetric method was based on the Dintenfass equation where blood viscosity, plasma viscosity and hematocrit are measured and used to calculate an index of RBC rigidity (Tk index). The RBC deformability/rigidity of the three groups was compared using the two methods. Tk index was not different between SS and SC patients and the two groups had higher values than AA group. When ektacytometry was used, RBC deformability was lower in SS and SC groups compared to the AA group and SS and SC patients were different. Although the two measures of RBC deformability were correlated, the association was not very high. Bland and Altman analysis demonstrated a 3.25 bias suggesting a slight difference between the two methods. In addition, the limit of agreement represented 28% (>15%) of the mean values of RBC deformability, showing no interchangeability between the two methods. In conclusion, measuring RBC deformability by indirect viscosimetry is less accurate than by ektacytometry, which is considered the gold standard.
Fichier non déposé

Dates et versions

hal-01881561 , version 1 (26-09-2018)

Identifiants

Citer

Jens Vent-Schmidt, Xavier Waltz, Aurélien Pichon, Marie-Dominique Hardy-Dessources, Marc Romana, et al.. Indirect viscosimetric method is less accurate than ektacytometry for the measurement of red blood cell deformability.. Clinical Hemorheology and Microcirculation, 2018, pp.115-21. ⟨10.3233/CH-131727⟩. ⟨hal-01881561⟩
77 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More